Multimetro Digital Ecrã Duplo 3 3/4 Digitos
Multi and large Display with 3¾ digit main display, two 9999 count subsidiary display and analog bar graph
RS-232C interface function for communicating with Personal Computer
TRUE RMS AC Voltage Measurement
REC function for displaying MIN. AVG. MAX. simultaneously
Comparison function for in-tolerance limits setting (REL/REL % function)
Menu-Driven operation for edit & shift
Go/No-Go Limit value editing
Overload protection on all range
Auto ranging and manual selection
Auto power off to save power
Battery use ; 9V 6F22 x 1pc
DC Voltage 400mV 100uV +/- (0.3% + 2d) 10 Mohm
4V 1mV
40V 10mV
400V 100mV
1000 V 1V
Range Resolution Accuracy Input resistance
45Hz-450Hz 450Hz-5kHz 5 - 20kHz
AC Voltage 4 V 1 mV +/-(0.75%+5d) Unspec,d Unspec,d 10 Mohm
40 V 10 mV +/-(2%+5d) +/-
(2.5+5d)
400 V 100 mV
750 V 1 V Unspec,d
Range Resolution Accuracy Over Load Protection
DC Current 40 mA 0.01 mA +/- (0.5% +5d) 0.5A / 250V
400mA 0.1 mA
4 A 0.001 A +/- (0.75% + 5d) 10A / 250V
10 A 0.01 A
AC Current 40 mA 0.01 mA +/- (0.75% + 5d) 0.5A / 250V
400 mA 0.1 mA
4 A 0.001 A +/- (1.5% + 5d) 10A / 250V
10 A 0.01 A
Resistance 400 ohm 0.1 ohm +/- (0.5% + 5d) 600 V DC or AC Peak
4 kohm 1 ohm
40 kohm 10 ohm
400kohm 100 ohm
4 Mohm 1 kohm
40 Mohm 10 kohm +/- (1% + 10d)
Frequency 200 Hz 0.01 Hz +/- (0.05% + 2d) 600 V DC or AC Peak
2 kHz 0.1 Hz
20 kHz 1 Hz
200 kHz 10 Hz
Capacitance 400 nF 0.1 nF +/- (3% + 5d) 600 V DC or AC Peak
4 uF 0.001 uF
40 uF 0.01 uF
Test Voltage Threshold Over Load Protection
Audible Continuity 1.5 V Less than 100ohm 250V DC or AC
Test Voltage Max. Test Current Over Load Protection
Diode Test 3 V Approx. 1mA 250V DC or AC
Baudrate Parity bit Data bit Stop bit Remark
RS-232C Interface 1200 None 7 1
General Test Condition
Power Requirement 9 V 6F22 x 1
Dimension 90(W) X 190(D) X 32(H) mm
Weight Approx. 750 g
Standard Accessories Test Lead, Instruction Manual, RS-232C software, Rs-232C Cable
Optional Accessories Rubber Holster